Hälso- och sjukvård
- +Ämnesområden
- +Hälso- och sjukvård (20)
- +Medicin: allmänt (1)
- +Medicinsk utrustning (7)
- +Tandvård (7)
- +Sterilisering (1)
- +Laboratoriemedicin (3)
- +Kvalitetsledningssystem inom hälso och sjukvård (0)
- +Farmakopéprodukter (1)
- Sjukvårdstextilier: allmänt (0)
- Första hjälpen-utrustning (0)
- +Hjälpmedel för personer med funktionsnedsättning (0)
- Födelsekontroll. Mekaniska preventivmedel (0)
- Veterinärmedicin (1)
This document specifies the characteristics of, and corresponding test methods for, unalloyed titanium for use in the manufacture of surgical implants. Six grades of titanium based on tensile strength are listed in Table 2. NOTE The mechanical properties of a sample obtained from a finished product made of this metal do not necessarily comply with those specified in this document.
This document specifies requirements and test methods for powered polymerization activators in the 380 nm to 515 nm wavelength region intended for chairside use in polymerization of dental polymer-based materials. This document applies to quartz-tungsten-halogen lamps and light-emitting diode (LED) lamps. Powered polymerization activators could have internal power supply (rechargeable battery powered) or be connected to external (mains) power supply. Lasers or plasma arc devices are not covered by this standard. This document does not cover powered polymerization activators used in laboratory fabrication of indirect restorations, veneers, dentures or other oral dental appliances.
This document specifies the requirements and test methods for sterilized absorbent points used in endodontic procedures. Absorbent points are marketed sterilized or non-sterilized. The requirements apply to absorbent points which have been sterilized once in a manner approved by the manufacturer. This document specifies numerical systems and a colour-coding system for designating the sizes of absorbent points.
Clause 7 specifies the labelling and packaging needed, including the instructions for use. A claim by the manufacturer that the contents of the unopened pack are sterile is the responsibility of the manufacturer (see Table 2). This document does not specify requirements or test methods for sterility.
NOTE 1 Reference to applicable national regulations can be made.
Reference is made to internationally accepted pharmacopoeia.
NOTE 2 National requirements can apply.
Standards on methods of validating sterilization processes are also available: ISO 11137-1, ISO 11137-2 and ISO 11137-3.
This document classifies polymer-based crown and veneering materials used in dentistry and specifies their requirements. It also specifies the test methods to be used to determine conformity to these requirements. This document is applicable to polymer-based crown and veneering materials for laboratory fabricated permanent veneers or crowns. It also applies to polymer-based dental crown and veneering materials for which the manufacturer claims adhesion to the substructure without macro-mechanical retention such as beads or wires.
This standard lists the terms and definitions to be used in the IDMP (Identification of medicinal products) standards and technical specifications, when terms and definitions as to be used in more than one of these standards.
ISO 10993-11:2017 specifies requirements and gives guidance on procedures to be followed in the evaluation of the potential for medical device materials to cause adverse systemic reactions.
ISO 10322-1:2016 specifies requirements for the optical and geometrical properties of all semi-finished single-vision and multifocal spectacle lens blanks.
ISO 11979-4:2008 specifies the labelling requirements for intraocular lenses (IOLs) and the information to be provided within or on the packaging.
ISO 13504:2012 specifies general requirements for the manufacture of instruments and related accessories used in the placement of dental implants and further manipulations of connecting parts in the craniofacial area. It is applicable to single-use and reusable instruments, regardless of whether they are manually driven or connected to a power-driven system.
This document specifies a process for a medical laboratory to identify and manage the risks to patients, laboratory workers and service providers that are associated with medical laboratory examinations. The process includes identifying, estimating, evaluating, controlling and monitoring the risks. The requirements of this document are applicable to all aspects of the examinations and services of a medical laboratory, including the pre-examination and post-examination aspects, examinations, accurate transmission of test results into the electronic medical record and other technical and management processes described in ISO 15189. This document does not specify acceptable levels of risk. This document does not apply to risks from post-examination clinical decisions made by healthcare providers. This document does not apply to the management of risks affecting medical laboratory enterprises that are addressed by ISO 31000, such as business, economic, legal, and regulatory risks.
This document specifies general requirements and test methods for portable dental equipment for use in non-permanent healthcare environments. Portable dental equipment within the scope of this document includes portable dental units, portable patient chairs, portable operator's stools, portable operating lights, portable suction source equipment, portable air compressors and other portable dental equipment in instances where these devices are designed and constructed to be transported for use in non-permanent healthcare environments. NOTE Particular requirements for specific types of portable dental equipment for use in non-permanent healthcare environments are specified in subsequent parts of this document. This document does not apply to stationary dental equipment, wearable equipment (such as headlamps and loupes), mobile dental equipment or portable dental equipment that is not intended to be used in non-permanent healthcare environments or not designed to be disassembled, folded or packed for human transport between non-permanent healthcare environments. Also, requirements for stationary dental equipment that can be installed in a dental mobile medical facility (e.g. vehicular or containerized mobile dental clinic) are not considered in this document.
ISO 10993-3:2014 specifies strategies for risk estimation and selection of hazard identification tests, with respect to the possibility of the following potentially irreversible biological effects arising as a result of exposure to medical devices: genotoxicity; carcinogenicity; reproductive and developmental toxicity. ISO 10993-3:2014 is applicable when the need to evaluate a medical device for potential genotoxicity, carcinogenicity, or reproductive toxicity has been established.
This document specifies requirements and test methods for operating lights used in the dental office and intended for illuminating the oral cavity of patients. It also contains specifications on the instructions for use, marking and packaging. This document applies to operating lights, irrespective of the technology of the light source. This document excludes auxiliary light sources, for example, from dental handpieces and dental headlamps and also operating lights which are specifically designed for use in oral surgery.
This document specifies requirements for fire activated oxygen shut-off devices that stop the flow of oxygen in respiratory therapy tubing when activated by fire. NOTE 1 Typical arrangements for fire activated oxygen shut-off devices are shown in Annex C. NOTE 2 Respiratory therapy tubing is covered by ISO 17256. NOTE 3 Use of fire activated oxygen shut-off devices in medical devices or accessories is not mandated in this document. The fire activated oxygen shut-off devices specified in this document are not suitable for use with oxygen therapy systems with flows in excess of 20 l/min). NOTE 4 There is rationale for this clause in A.2. The requirements in this device-specific standard take precedence over any conflicting requirements in the general standard for airway devices (ISO 18190). All the common requirements that appear in the general standard for airway devices have been removed from this document.
This document specifies the control and approval of in vitro diagnostic reagents used in animal health for the detection, and/or absolute quantification of pathogen-specific nucleic acid (DNA or RNA) by PCR (e.g. endpoint PCR, real-time PCR, reverse transcription-PCR). This document is applicable to diagnostic reagents as a priority for infectious diseases (due to bacteria, viruses, fungi, or parasites, including genetic markers associated with pathogenicity, such as antimicrobial resistance or toxin production) and associated animal species for which harmonization of practices in this area is needed, i.e. those for which the national, regional or international regulatory framework provides for the control of trade in animals and/or animal products and/or the definition of a health status (absence of infection) of areas, establishments or individuals. Anyhow, all reagents designated by the competent authorities fall under the scope of this document. Nevertheless, the authorities or any other animal health stakeholder can choose to derogate in specific and very limited situations such as emerging, exotic or rare diseases. This document is not applicable to all existing diagnostic reagents, in particular those for which certain parameters described in this document cannot be validly evaluated in accordance with international requirements, due, e.g. to the absence of a specific reference standard and/or accessible and duly validated reference materials. The PCR diagnosis usually involves the use of a nucleic acid extraction and/or purification reagent, and a PCR reagent. The PCR method (when applicable) involves the successive use of these distinct reagents. PCR reagent control can be performed if the applicant provides evidence of the validity of the PCR reagent for use in the animal health diagnostic analysis, by proving its diagnostic performances with nucleic acid extracts obtained from the different matrices described in the instruction for use. The control of a complete PCR method by the applicant and the control organization is performed only if the PCR reagent cannot be dissociated from an nucleic acid extraction and/or purification systems. This document does not cover the control of the nucleic acid extraction and/or purification reagents, only. This document does not cover the step in which the user verifies a reagent (analysis method adoption). NOTE Prion diseases are not included in the scope of this third part of the EN 18000 series. Unlike other infectious diseases, prion diseases are not diagnosed using PCR assays because prions lack a nucleic acid component and consist solely of an abnormally folded conformer of the normal host protein.
Detta dokument avser regelbunden rengöring av lokaler och inventarier inom hälso- och sjukvård för att minska smittspridning.
Detta dokument anger:
kriterier för olika hygien- och kvalitetsnivåer
krav på och rekommendationer för uppföljning och kontroll av rengöring
krav på utbildning och kompetens hos personalen samt ansvarsförhållanden
krav på städutrustning samt
rekommendationer för dokumenterade instruktioner om hur rengöringen utförs
rekommendationer för städfrekvens
Detta dokument omfattar inte rengöring som regleras i specifika lagar och föreskrifter.
Exempel: Vissa lokalfunktioner regleras i andra dokument såsom tex Good Manufacturing Practice (GMP).
ISO 8980-4:2006 specifies optical and non optical requirements, including durability, and test methods for anti-reflective coatings on spectacle lenses. ISO 8980-4:2006 does not deal with the following topics: transmittance and absorbtance; the colour of the reflected light.
ISO 18739:2016 specifies terms, synonyms for terms and definitions used in the process chain for CAD/CAM systems in dentistry.
The base nomenclature provided in IEEE 11073 to support terminology for implantable cardiac devices is extended in this standard. Devices within the scope of this nomenclature are implantable devices such as pacemakers, defibrillators, devices for cardiac resynchronization therapy, and implantable cardiac monitors. The discrete terms necessary to convey a clinically relevant summary of the information obtained during a device interrogation are defined in this nomenclature. To improve workflow efficiencies, cardiology and electrophysiology practices require the management of summary interrogation information from all vendor devices and systems in a central system such as an Electronic Health Records (EHR) system or a device clinic management system. To address this requirement, the Implantable Device, Cardiac (IDC) Nomenclature defines a standard-based terminology for device data. The nomenclature facilitates the transfer of data from the vendor proprietary systems to the clinic EHR or device clinic management system.