Detaljer om förslaget

Svensk titel: Hydraulik - Kalibrering av automatiska partikelräknare för vätskor (ISO 11171:2022, IDT)
Nummer:ISO 11171:2022
Källa:ISO
Kommitté:SIS/TK 108
Namn på kommittén:Renhetsteknik
Publicerat:den 29 okt 2024
Sista svarsdatum:den 28 dec 2024
Contact email:kommentera(at)sis.se
Förslagets omfattning:

This document specifies procedures for the following:

a) primary particle-sizing calibration for particle sizes 1 µm(c) and larger, sensor resolution and counting performance of liquid automatic particle counters that are capable of analysing bottle samples;

b) secondary particle-sizing calibration using suspensions verified with a primary calibrated APC;

c) establishing acceptable operation and performance limits;

d) verifying particle sensor performance using a test dust;

e) determining coincidence and flow rate limits.

This document is applicable for use with hydraulic fluids, aviation and diesel fuels, engine oil and other petroleum-based fluids. This document is not applicable to particle-sizing calibration using NIST SRM 2806b primary calibration suspensions.

Detta dokument föreslås att fastställas och publiceras som svensk standard.

För att lämna allmänna kommentarer på standardförslaget i sin helhet och på föreslagen titel – klicka på förslagets titel.

För att lämna specifika kommentarer på innehållet – skriv kommentaren och föreslagen ändring i rutan under avsnittet i fråga.

Standardförslag på engelska föreslås att fastställas som svensk standard utan översättning. Kommentarer på engelskspråkiga standardförslag bör vara på engelska.

För att ett standardförslag ska kunna fastställas som svensk standard måste det vara förenligt med svensk lagstiftning. Det är således viktigt att berörda svenska myndigheter klarlägger ifall standardförslaget är förenligt med svensk lagstiftning.