Detaljer om förslaget

Nummer:prEN 12543-5
Källa:CEN
Kommitté:SIS/TK 125
Namn på kommittén:Oförstörande provning
Publicerat:den 19 jul 2024
Sista svarsdatum:den 16 sep 2024
Contact email:kommentera(at)sis.se
Förslagets omfattning:

This document specifies a method for the measurement of focal spot dimensions within the range of 5 µm to 300 µm of X-ray systems up to and including 225 kV tube voltage. This determination is based on the evaluation of an image with a dedicated focal spot that has been radiographically recorded using an edge and evaluated with a digital method. The imaging quality and the resolution of X-ray images depend highly on the characteristics of the effective focal spot, in particular the size and the two-dimensional intensity distribution as seen from the detector plane. For the characterization of commercial X-ray tube types (i.e. for advertising or trade) the specific maximum values of Annex A are preferred. NOTE The same procedure can be used at higher kilovoltages by agreement but the accuracy of the measurement can be poorer.

Detta dokument föreslås att fastställas och publiceras som svensk standard.

För att lämna allmänna kommentarer på standardförslaget i sin helhet och på föreslagen titel – klicka på förslagets titel.

För att lämna specifika kommentarer på innehållet – skriv kommentaren och föreslagen ändring i rutan under avsnittet i fråga.

Standardförslag på engelska föreslås att fastställas som svensk standard utan översättning. Kommentarer på engelskspråkiga standardförslag bör vara på engelska.

För att ett standardförslag ska kunna fastställas som svensk standard måste det vara förenligt med svensk lagstiftning. Det är således viktigt att berörda svenska myndigheter klarlägger ifall standardförslaget är förenligt med svensk lagstiftning.